1 Maintenance and repair

The Sensistor XRS9012 is maintenance

free, except for battery replacement (normal-

ly every three years). This section describes

the functions of the instrument and gives

advice as how to function tests should be

carried out.

1.1 Function and controls

Following functions are controlled from the

panel:

- Sensitivity: can be varied in 10 steps, each

step doubles the sensitivity.

- On/Off: the detector has an automatic

switch-off aftr 30 minutes of no activity.

- Battery Check: At least 8 of the light emit-

ting diods (LED) will illuminate at full

charge. Full charge takes one hour.

- Zero: the detector has an automatic zero-

ing function. The ZERO button allows the

operator to force the signal to zero instanta-

neously.

- Sound level: the volume can be changed in

64 steps.

- Earphone output

- MAX: this is a peak-hold function allow-

ing the operator to compare the maximum

gas signal in two different places.

- SEARCH (default): this is the normal

mode where the signal follows the gas con-

centration. The detector always starts in this