1.4 Spare part list

Part no.

Battery

591-294

Earphones 591-443

Carrying case

591-304

Charger AC

(591-300) 591-301 SV, 591-302 USA, 591-303 UK

Charger car

591-361

Hand Probe H21

590-200

Cable C21

590-150

Circuit board

591-293

1.5 Function check

Action

Result

Connect charger The instrument starts

Disconnect The instruments turns off

Change sensitivety The LED’s indicates corresponding sensitivity

Adjust the volume The volume should change

Press “Battery Check” button The LED should indicate Battery status

Disconnect Hand Probe The ON LED should flash red and an audio

signal appear

1.6 Sensitivity check

Action

Result

Start XRS9012 The instruments starts and indicates sensitivity 5

Adjust the sensistivity to 10 The no. 10 LED is flashing

Apply 10ppm Hydrogen Signal should indicate: minimum LED 1