Hydrogen leak detector Sensistor ISH2000 HySpeed

-

Operating Instructions


OP-Sensistor ISH2000 HySpeed-EN-201203

37

EN

Table 8-11. Common used functions.

K = Calibration request

Starts calibration if the Sensistor ISH2000 HySpeed has an active probe driver
installed. Sensistor ISH2000 HySpeed answers with a "K" if an active driver
containing a calibration routine was found and "F" if the calibration APC sequence
not found. Calibration doesn’t start if Purge level is reached.

M = Measure Request

The active test cycle defined by the APC driver starts. “M” is returned if the selected
driver supports active test. “F” (failed) is returned otherwise.

N = Print Request

Returns current analysis value.

Q = Set APC in stand by, (stop a measurement)

Returns a “Q”.

R = Activates probe 0 (built in P50/Combox with H65 driver)

Returns an “R”.

S = Activates probe 1 (Installed probe driver or trigg signal handler driver)

Returns an “S”.

X = Shift State to "Analysis Mode"

Returns nothing.

Z = Shift State to "Detection Mode"

Returns nothing.

Y = Shift State to “Combined Mode”

Return nothing.

Supported parameters

Table 8-12. The following parameters can be downloaded to Sensistor ISH2000
HySpeed in

Analysis

and

Detection mode

.

Command

Header

Calibrate

K

Measure M

Print Request

N

Stop Measurement

Q

Hand Probe

R

Active Probe (Installed AP)

S

Analysis Mode

X

Detection Mode

Z

Combined Mode

Y

Parameter

Header

Data

Reject Level

A

n.nnE+nn

Correlation Value

B

n.nnE+nn

Analysis Unit

C

Text string (max 12 characters)

Analysis Unit

CUx

x=1 to 8, 1=PPM, 2=CC/S up to 8