beautypg.com

Fig 8-5. d9 null modem cable wiring diagram, Selectable printer types, Table 8-5. communication specifications – INFICON Sensistor ISH2000 HySpeed Hydrogen Leak Detector User Manual

Page 34

background image

Hydrogen leak detector Sensistor ISH2000 HySpeed

-

Operating Instructions

34

OP-Sensistor ISH2000 HySpeed-EN-201203

EN

Only pin 2 (Received data), pin 3 (Transmitted data) and pin 5 (Signal ground) are
used. Refer to Figure 8-5 for the wiring diagram.

Fig 8-5. D9 Null modem cable wiring diagram

Selectable printer types

Most PC-printers with serial interface can be connected to the 9-pin printer port.
Parallel (Centronics) interface printer can be used if connected through a serial to
parallel converter.

The port can be set up for the following printer types: PC Printer and Data Dump.

No printer

Printer output disabled. Incoming communication is enabled. Sensistor ISH2000
HySpeed listens for incoming data but will not print/send test results.

PC printer (with serial interface)

The

PC Printer

option can be used to print data on most standard PC printer with

serial interface. Parallel interface printers can be used if connected through a serial
to parallel converter (see below).

Note:

The output format has been chosen to be as simple as possible to ensure that
most printers will accept it. Therefore, the printer output does not use any
flow control. This means that some printers may delay printing until the input
buffer is full or a pre-defined timeout has elapsed.

Table 8-5. Communication specifications.

Note:

Due to the large variety of printers available on the market, INFICON does not
take responsibility for the operation of a particular type of printer.

6

(DSR)

Not used

7

(RTS)

Not used

8

(CTS)

Not used

9

(CE)

Not used

Pin

Signal

Comments

1

2

3
4

5

6

7
8

9

9

8

7

6

5

4

3

2

1

CD

CD

RxD

RxD

TxD

TxD

DTR

DTR

SG

SG

DSR

DSR

RTS

RTS

CTS

CTS

R1

R1

D9 Female

D9 Female

Setting

Value

Data rate

1200 baud

Data bits

8

Stop bits

1

Parity

None

Flow control

None