9 general settings, General settings – INFICON Sensistor ISH2000P Hydrogen Leak Detector, Panel Model User Manual
Page 29

Hydrogen leak detector Sensistor ISH2000
-
Operating Instructions
OP-Sensistor ISH2000-EN-201203
27
EN
minutes, the function is deactivated if set to OFF. The display lamp will return to the
set brightness if any of the display buttons are pressed, if a gas signal is detected or
an instrument error is detected.
8.9 General settings
This part describes the general settings of the Sensistor ISH2000.
Language
The Sensistor ISH2000 user interface contains the following languages:
•
English
•
French
•
German
•
Italian
•
Spanish
•
Swedish
Measure/Print Button
Setting this parameter to
ON
displays
Measure
or
above the button 1.
Measure
will be displayed for an APC-Probe or
for a Hand Probe. Pressing
Measure
will
initiate a sample cycle. Pressing
will send the values from the hand probe
measurement to the printer port.
Probe Button
This is for setting the different functions with the probe button. These functions are
as follows:
•
Toggle Mode-makes it possible to switch between Analysis mode and Detection
mode.
•
Zero detection signal-in Analysis mode and Detection mode.
•
Measure/Print-makes it possible to initiate sample cycles or send the values from
the hand probe measurement to the printer port.
•
Probe Lamp-makes it possible to turn on and off the Probe Lamp.
Probe Lamp
Makes it possible to have the Probe Lamp on even if the other Probe Button
function is chosen.
Change Password
The user password is a text string (max 12 alphanumerical characters) used to lock
critical parameters. Setting password to an empty string (no characters) means that
no password is needed to modify the critical parameters. The default is no password
(“”).
Contact INFICON AB if you have lost your user password. If the
Password Protected
Calibration
parameter is set to
ON
you will be prompted for a password when starting
a calibration.
Note:
Setting
Password Protected Calibration
to
ON
has no effect if no password is set.
Note:
APC controlled calibration can be started from the bus in both cases.