Install collecting flask in cassette – Heidolph LABOROTA 20 compact and safety User Manual
Page 73
73
E
7. Install collecting flask in cassette
Loosen and remove 4 ea. knurled knobs (1).
Lift and remove top panel (2).
Install collecting flask (6) in bottom panel (3).
Aerating fitting (4) needs to point forward.
Reinstall top panel (2) and secure with the 4 knurled
knobs (1)
Install cassette assembly. (with collecting flask) on
bracket (8)
Hold handle (7) of bracket (3), loosen thumbscrew
(5), and lower bracket completely; install cassette
completely (direction as marked).
Lift handle (7) until ground joint of collecting flask
reaches mating end of condenser, secure with
thumbscrew (5)
use KS45 flask clamp to secure both parts in place
8
See also other documents in the category Heidolph Equipment:
- Hei-VAP Precision (59 pages)
- Hei-VAP Value (84 pages)
- Rotacool Mini (104 pages)
- Rotacool (15 pages)
- RotaChill (37 pages)
- ROTAVAC valve control (188 pages)
- ROTAVAC valve control (174 pages)
- ROTAVAC vario control (34 pages)
- ROTAVAC Vario pumping unit (61 pages)
- LABOROTA 4000 eco (168 pages)
- LABOROTA 4000 up to 4003 (364 pages)
- Heating bath HB 4000 (52 pages)
- LABOROTA 20 eco (180 pages)
- LABOROTA 20 automatic (190 pages)
- ROTAVAC 20 (35 pages)
- Automatic Module Distimatic (52 pages)
- Synthesis 1 (268 pages)
- MR Hei-Standard (39 pages)
- MR Hei-End (124 pages)
- MR Hei-Mix L (24 pages)
- MR Hei-Mix S (34 pages)
- EKT Hei-Con (78 pages)
- Heat-On (13 pages)
- StarFish Multi-Experiment Work Station (17 pages)
- RZR 1 (44 pages)
- RZR 2020 (58 pages)
- RZR 2051 control (68 pages)
- REAX top (44 pages)
- REAX 2 (48 pages)
- REAX 20 (54 pages)
- Multi Reax (60 pages)
- TITRAMAX 100 / 101 / 1000 (60 pages)
- UNIMAX 1010 (68 pages)
- UNIMAX 2010 (72 pages)
- Incubator 1000 (64 pages)
- Pump drives (198 pages)
- SilentCrusher M (146 pages)
- SilentCrusher S (98 pages)