Chapter 5 error indication, Chapter 4 setting parameters – DIGIWEIGH DWP-11KRW User Manual
Page 11
Floor / Pallet Scale
- 10 -
Chapter 4 Setting Parameters
To enter parameter setting: Turn on the indicator and press [ZERO] key when the indicator is still
warming up.
4.1 “F1 CLK” DATE AND TIME SETTING
Press [FUN] key to show the current time. If it is not correct, adjust it as following:
Press [ZERO] key to change hour (the hour is blinking). Press [TARE] to add up until it shows the
correct hour. Press [ZERO] to move to the minutes (the minutes are blinking). Press [TARE] to add up
until it shows the correct minute. Press [ZERO] to move to the seconds (the seconds are blinking).
Press [TARE] to add up until it shows the correct second.
Press [FUN] to confirm and move to date mode setting. There are 3 date showing modes:
“YY-MM-DD”, “MM-DD-YY” and “DD-MM-YY”. Press [TARE] key to switch between and press [FUN]
to confirm then move to date setting. Use the same way as timing setting to set the date. Press
[FUN] to confirm and exit. During the time or date setting, you may press [MOD] anytime to cancel
and exit. System will not save the changes. Press [TARE] to move to the next parameter.
4.2 FAST CALIBRATION FOR ZERO POINT AND FULL RANGE POINT
Press [#] while it is initialization, it enters into the calibration mode.
4.2.1 Fast calibration for zero point:
At any time before it shows [nOLOAD], press [FUNC], it keeps the original division, decimal point, full
range and enter into the zero point calibration mode. Press [ZERO] when the stable signal is on, it
displays [End] and keeps the original full range point calibration. Press the calibration switch under
the lead sealing board at the back of the indicator, it saves the setting and back to the weighing status.
4.2.2 Fast calibration for full range point:
At any time before it shows [AdLOAD], press [ACCU], it keeps the original division, decimal point, full
range, zero point calibration and enter into the full range point calibration mode. When it is finished,
press the calibration switch under the lead sealing board at the back of the indicator, it saves the
setting and back to the weighing status.
Chapter 5 Error Indication
EER 1
The AD value is too small when calibrated.
EER 2
The zero point is out of range when calibrated.
EER 3
The zero point is out or range upon starting
EER 4
The imputed sample number is zero when sampling n counting mode.
EER 5
The imputed weight is zero when full scale calibrated in calibrating mode.
EER 6
The unit weight is less than 0.25e when sampling in counting mode
bAt-lo
Low power