Optional mrc5000 recorder setup – Despatch LAD Series WITH PROTOCOL PLUS User Manual
Page 68

60
Optional MRC5000 Recorder Setup
The temperature is retransmitted to the Recorder from the Controller. Set up the
Recorder as follows:
1.
Make sure that jumper JU1 is setup for the 5 VDC setting (see MRC Manual).
2.
Move the Mode switch to the PROG/TEST/CAL position, and Prog will be
displayed.
3.
Press the down arrow key twice and Inps will be displayed. Make sure the
settings are per the table below.
4.
Once all the settings have been changed, move the Mode switch to the RUN
position. The display on both the Recorder and controller should read the same.
Parameter Code
Degrees C
Degrees F
Inps
17
17
Icor
0
0
diSP
On
On
dPOS
0
0
EUU **
400
752
EUL **
0
32
ChUP
400
800 *
ChLO
0
0
DFF
1
1
* Note:
The 0 - 400 chart paper must be changed to the 0 - 800 chart paper.
Depending on the equipment, 0 - 600 paper can be used if the maximum
temperature is 500 degrees F.
** Note:
These values must mach the setting sRetOutLo and RetOutHi on the
Control page on the Protocol Plus controller (example: RetOutLo is 32,
then EUL must read 32).