Despatch 926E Equipment Manual (C-155)Rev 2/11 User Manual

Page 16

background image


11

Pressure of Solid, Liquid and Saturated Vapor CO

2

Temperature (

F)

Pressure (PSIA)

Temperature (

F)

Pressure (PSIA)

-140

3.19

-20

215.02

-130

5.39

-10

257.46

-120

8.85

0

305.76

-110

14.22

+10

360.4

-100

22.34

20

421.8

-90

34.05

30

490.6

-80

50.70

40

567.3

-70

74.90

50

652.7

-69.9

75.1

(triple point)

60

747.4

-60

94.75

70

852.5

-50

118.27

80

969.3

-40

145.87

87.8

1072.1

-30

177.97


The solenoid valve used to inject the CO

2

coolant into the test chamber should "pulse" or

actuate rapidly. If the CO

2

valve remains open for long periods of time after pull-down, the

chamber may be leaking coolant, the test object may be introducing substantial amounts of
heat beyond the capacity of the inlet valve orifice, or the temperature set-point may be set
below -73.3

C (-100

F). These conditions can cause "snow" to be introduced into the

chamber with resultant control problems and possible damage to the chamber and contents.
Causes for these conditions should be located and corrected.