beautypg.com

Despatch 926E Equipment Manual (C-155)Rev 2/11 User Manual

Page 15

background image


10

A filter installed in the line must be checked for cleanliness every 50 hours of

operation, or more often if necessary. Clean the filter in trichloroethylene and
dry thoroughly using filtered, moisture-free compressed air at low pressure.


It is important that the CO

2

solenoid valve assembly and the orifice be kept clean.



Additional Precautions

Never install CO

2

supply pipe fittings or valves of a larger internal diameter

than those used upstream [beginning at the supply cylinder(s)].

Do not use fittings which allow undue expansion or restriction of the liquid

CO

2

. This could cause the CO

2

to reach a critical point where it would

expand in the line and cause icing.

Do not use excessively long, small diameter supply hoses which might cause

the liquid to change to gas in hot areas or during intermittent cycling
conditions.

Do not meter the CO

2

through any type of valve that has a smaller I.D. than

the supply hose and CO

2

supply cylinder siphon tube. The cylinder supply

valve should be fully open during use.

It is not possible to use all of the CO

2

in a siphon-type cylinder. The ambient

temperature at the cylinder will determine the amount of CO

2

available (about

95 percent of maximum). Thus, CO

2

usage may appear to be somewhat

higher than it actually is.

Never set the temperature controller below -100

F (-73

C). Excessive

snowing occurs as you approach temperature below -100

F (-73

C) until the

temperature reaches -109

F (-78

C) which is the freezing point of CO

2

at

atmospheric pressure. At -109

F (-78

C) all discharged CO

2

could form

snow which could cause damage to the chamber and its contents and if
allowed to build up to the point of blocking off the pressure relief vent, could
cause explosion of the chamber due to available pressure from the liquid
carbon dioxide source.

This manual is related to the following products: