Enhanced mode – Carbolite 302P4 User Manual
Page 21

Programming
21
MC22-GB-V1.00
5
Enhanced Mode
In enhanced mode the following parameters are available.
CUSTOMER CALIBRATION TYPE SELECTION
Identification code: CL.st
Range:
FACt
Factory calibration
(a = 1 b = 0)
C.CL1
1 point calibration
(a = 1 b = OFSt)
C.CL2
2 point calibration
(a = x b = x)
C.CL3
2 coefficients calibration
(a = SLOP b = OFSt)
OFFSET
(This parameter is always available but can be modified only if CL.ST = C.CL1, C.CL3)
Identification code: OFSt
Range:
-199.9 / 199.9
SLOPE
(This parameter is always available but can be modified only if CL.ST = C.CL3)
Identification code: SLOP
Range:
0.500 / 2.000
LOW POINT CALIBRATION
(This parameter is available only if CL.ST = C.CL2)
Identification code: CaL.L
Range:
Value in engineering units between min / max range value
OFFSET ON LOW POINT
(This parameter is available only if CL.ST = C.CL2)
Identification code: OFS.L
Range:
-199.9 / 199.9
HIGH POINT CALIBRATION
(This parameter is available only if CL.ST = C.CL2)
Identification code: CaL.H
Range:
Value in engineering units between min / max range value
OFFSET ON HIGH POINT
(This parameter is available only if CL.ST = C.CL2)
Identification code: OFS.H
Range:
-199.9 / 199.9
MODBUS ADDRESS
Identification code: adr
Range:
0 (Interface disable) 1 up to 254