Programming
12
MC22 -GB-v1.01
itSp
Meaning:
initial set point, it is the set point used by the programmer during the
“idle” state
Range:
OFF, and then from 0°C – 1500°
The OFF value means that the control output(s) will be disabled
during the “idle” state; at program start up the setpoint will be
aligned to measure value and then ramps to SP. 1.
Available:
always
Alterable:
only when the programmer state is idle or, for program 1-2-3, if the edit is
enabled
Sp. 1
Meaning:
target set point of the first segment
Range:
0°C
– 1500°C
Available:
always
Alterable:
only when the programmer state is idle and, for program 1-2-3, the edit is
enabled
rr. 1
Meaning:
ramp rate to SP1.
Range:
from 0.1 to 12°C (Engineering units (°C) per minute)
Available:
always
Alterable:
only when the programmer state is idle and, for program 1-2-3, the edit is
enabled
TH. 1
Meaning:
time held (dwell) at SP1
Range:
from 0 min to 99 h 59 min
Available:
always
Alterable:
only when the programmer state is idle and, for program 1-2-3, the edit is
enabled
Sp. 2
Meaning:
target set point of the second segment
Range:
End, and then from 0°C – 1500°C
The End value, selectable from the second to the fourth set point, is used to
end the program before the last
– fourth - segment. This avoid to edit
unnecessary parameters for programs shorter than four segments.
Available:
always
Alterable:
only when the programmer state is idle and, for program 1-2-3, the edit is
enabled
Then rr. 2 , tH. 2, Sp. 3 , rr. 3 , tH. 3 , Sp. 4 , rr. 4 , tH. 4 as above.