MF47-3.05
19
9.0 S
PECIFICATIONS
Carbolite reserves the right to change specifications without notice.
9.1 Models Covered by this Manual
MODEL
Max.
Temp.
Max.
Power*
Chamber Size (mm)
Approx.
Capacity
Max.
Load
Net
Weight
(°C)
(kW)
H
W
D
(l)
(kg)
(kg)
HTF 17/5
1700°C
4.3
158
150
225
5.3
2.5
87
HTF 17/10
1700°C
6
232
200
225
10.4
5
106
HTF 18/4
1800°C
4.6
141
140
205
4
2
92
HTF 18/8
1800°C
6.2
210
190
205
8.2
4
113
*Maximum power as measured, included ancillary components and transformer losses
9.2 Environment
The furnaces contain electrical parts and should be stored and used in indoor conditions as follows:
temperature:
5°C - 40°C
relative humidity:
maximum 80% up to 31°C decreasing linearly to 50% at 40°C
altitude:
not exceeding 2000m
electrical supply:
fluctuation not exceeding 10%
overvoltage:
category II IEC60364-4-443
pollution:
degree 2