Carbolite PTC 12/20/150 User Manual
Page 9
PTC
MF11 – 3.08
9
4.8 Setting the Timer Mode
Scroll to tm.OP; use or to view and alter the mode. The mode shows as OPt.1 to OPt.5.
It is not possible to alter the mode while the timer is running; if the mode cannot be altered, scroll
to the
StAt
parameter and set its value to
OFF
.
4.9 Setting the Time Period
Method 1
Scroll to
tmr
(time remaining). Use or to view the remaining time; the units are always in
minutes. Use and to set or alter the time. Setting
tmr
automatically activates the timer; the
m-A
parameter changes to
Auto
and the
StAt
parameter changes to run.
Note that the
tmr
display shows
0
(zero) during the last minute of timing, and also shows
0
when
the time has expired. The timer light indicates whether timing is still in progress.
Method 2
Scroll to
dwEl
, and use and to set the timing duration. The advantage of method 2 is that
dwEl
need only be set once if repeated use of the same time period is required.
Scroll to
StAt
, and use or to set the parameter value to
run
. This copies the dwell time into
tmr
and activates the timer as in method 1.
4.10 Running with the Timer
Once the timer is activated by method 1 or 2 above, the control sequence depends on the timer
ode, as previously given in the table. Fig 3 gives another representation of the timer action.