Carbolite TLD Series User Manual
Page 8
TLD
8
MF18 – 3.06
5.6 Thermocouple Replacement
Remove the front panel and the internal cover as necessary.
Make a note of the thermocouple connections, and how the thermocouple is placed and fixed.
Colour codings are:
negative
positive (type K)
white
green
Disconnect the thermocouple from its controller terminals or terminal block.
Re-assemble with the new thermocouple observing the colour coding.
5.7 Element Replacement
Remove front panel and the internal cover as necessary and locate the element terminals.
Disconnect the wires from the element terminals. Remove any starlock washers - these may need
to be cut with wire cutters. Remove any clips holding the element inside the chamber, and
withdraw the element.
Reverse the procedure with the new element.
Run the oven at a low temperature and check that it is controlling properly, to find out whether the
element failure was caused by a fault in the control circuit.
5.8 Fuse Replacement
Fuses are accessed by removal of the panel near the cable entry point.
Depending on the model, supply fuses and control circuit fuses may be mounted in their own
holders, or may be on a circuit board which contains an EMC filter. The fuses are marked with
their ratings.
Take care not to disconnect the wires leading from the EMC filter without first recording their
positions: they must be reconnected to the correct terminals.