Installation – Carbolite TLD Series User Manual
Page 3
![background image](/manuals/555930/3/background.png)
TLD
MF18 – 3.06
3
2.0 I
NSTALLATION
2.1 Unpacking & Handling
Lift the unit by its base. Do not use the door or any other projecting cover or component to support
the equipment when moving it. Use two people to carry the oven where possible. Remove any
packing material from the inner chamber before use.
NOTE: This product contains Refractory Ceramic Fibre (better described as Alumino Silicate
Wool) for precautions and advice in handling this material see the ‘Repairs and Replacements’
section.
2.2 Siting & Setting Up
Place the oven on a level surface.
Ensure that there is free space around the oven. Ensure that any vents in the oven are not obscured.
During the operation of the oven hot air is exhausted from the vent on the top of the oven while
the cooling blower is in operation; there must be clearance of at least 500mm between the top of
the oven and any surface which must be made of non-combustible material.
Ensure that the oven is placed in such a way that it can be quickly switched off or disconnected
from the electrical supply - see below the top of the oven and the ceiling of the room. Always
ensure that the vent on the top of the oven is free to open.
2.3 Electrical Connections
Connection by a qualified electrician is recommended.
The oven is manufactured for use on a single phase supply, which may be Live to Neutral non-
reversible, Live to Neutral reversible or Live to Live.
Check the Oven rating label before connection. The supply voltage should agree with the voltage
on the label, and the supply capacity should be sufficient for the amperage on the label.
The supply should be fused at the next size equal to or higher than the amperage on the label. A
table of the most common fuse ratings is also given in section 8.1 of this manual. Where a supply
cable is present there are internal supply fuses; customer fusing is preferred but not essential.
Oven with supply cable: either wire directly to an isolator or fit with a line plug.
Oven without supply cable: a permanent connection to a fused and isolated supply should be made
to the internal terminals after temporary removal of the oven back panel.
Connection by line plug: the plug should be within reach of the operator, and should be quickly
removable.
Connection to isolating switch: this should operate on both conductors and should be within reach
of the operator.
The supply MUST incorporate an earth (ground).
CONNECTION DETAILS
supply type
Supply
Terminal label Cable colour
Live-Neutral
Reversible or Live-Live
1-phase
L
Brown
To live
to either power conductor
N
Blue
To neutral
to the other power conductor
PE
Green/Yellow
To earth (ground)
to earth (ground)