1 introduction and features, 2 specifications and constructions – C.B.S. Scientific DCX-800 User Manual
Page 5

5
www.cbsscientific.com
SECTION 1
General Information
1.1
Introduction and Features
C.B.S. Scientific offers the Wide Dual Mini-Verticals are for performing SDS-Page,
acrylamide-nucleic acid separations and electro-blotting. The dual unit provides the
capability of running or blotting two gels simultaneously. Units include 2 blotting cas-
settes, 2 foam pads, and a white plastic reservoir conversion plate to allow for single
runs. The DCX-800C has a cooling base for temperature controlled runs.
This system is designed to be used with ClearPAGE™ precast gels. (See Table 1).
Hand-cast gels can also be run in this unit with the use of accessory combs, plates and
spacers.
Table 1: Features of Vertical Mini-Gel Systems
1.2
Specifications and Constructions
Constructions:
Buffer chamber, safety cover, core
Polycarbonate/Polysulphone
Electrodes
Pure Platinum wire .010” diameter
Power cords
Silicone rated 7500VDC,
200mA,
65°C
Combs
Teflon
Glass plates
Soda-lime float glass & Borosilicate
Spacers
PVC
Clamps
Polypropylene,
stainless
steel
Safety Certification
EN61010-1-1993 (IEC1010-1)
Table 2: Specifications
Model #
DCX-800, DCX-800C
Shipping Weight
6 lbs
Overall Size W x D x H (cm)
22cm x 12.7cm x 15.4cm
Recommended buffer volumes:
Cathode reservoir
Anode reservoir
Total buffer volume for 2 blots
250mls
600mls (minimum) / 1100mls (maximum)
1.35 Liters
Slab gel run time
30-90 minutes
Blot run time
60-90 minutes
Handcast glass plate size (W x H)
15cm x 10cm
Voltage limit
300 V
Model #
Single or
Dual
Plate Dimensions
(w x h)
Compatible
Precast Gels
Cooling
DCX-800
DCX-800C
Dual
15cm x 10cm
ClearPAGE™ - 15cm(w)
NuPAGE Midi-Gels
Cooling base (in DCX-
800C only)