C.B.S. Scientific DTSK-2401-220 User Manual
Page 9

Cipher Genetic Analysis System
DTSK Instruction Manual, version 8/30/2011
9
www.cbsscientific.com
3.1 Unit Set-up and Unpacking Instructions-continued
4. Position an appropriate power supply on a shelf above the tank.
5. Fill the tank with running buffer. To obtain the correct level of buffer, place the cassettes in the tank and fill
the tank until the level of buffer reaches the underside of the upper reservoir. The upper reservoir holds only
50mls.
6. Attach the power supply to the safety interlock. (Fig 3-2-
A)
7. Attach the safety interlock to a wall (mains) outlet. (Fig 3-2-
B)
8. Plug the Heater/Stirrer into a wall (mains) outlet.
Important: Do not plug the heater/stirrer into anything other
than a wall outlet.
9. Do NOT attach power leads (from power supply to safety interlock, Fig 3-2-
C) until gels are loaded and
installed in the tank.
Power Supply/Safety Interlock Connections
Fig. 3-2
A
B
C