C.B.S. Scientific DTSK-2401-220 User Manual
Page 37

Cipher Genetic Analysis System
DTSK Instruction Manual, version 8/30/2011
37
www.cbsscientific.com
4. Place the notched plate on top of the bottom assembly, starting from
the bottom edge and gently easing the plate down. Verify the gasket is
smooth around the edges and then clamp along the bottom.
5. Lift the assembly and stand it on the base of the clamps. For leveling,
push glass plate assembly down until it stops against clamp body.
Clamp the sides of the assembly with additional casting clamps on
either side. As each clamp is attached, be sure the gasket is aligned
between the plates forming a seal. The glass plate assembly is ready to
receive the acrylamide solution, sec. 3.7.
6. Go to section 3.8 and follow instructions under the heading “Preparation of the Cassettes”.