Accessing the Mode Setting:

Step 1: Simultaneously press and hold the On/Off button and the (S) button

for approximately 3 seconds. This will bring up the 3 digit mode
number.

Step 2: Enter the desired mode number (see mode definitions below)

by using (Z) to change the value of the flashing digit, and (S) to
advance to the next digit.

Mode Definitions:

Mode 000: Full Access

- Provides access to all menu settings. If no buttons are pushed for 1
minute, the gauge will revert back to Mode 003 (factory default).

Mode 002: Limited Access

- Menu is locked
- (Z) button zeros the reading
- (S) cycles between the Max and Min readings
- On/Off functions only on battery powered gauges

Mode 003: Factory Default

- Full access except Full Scale Calibration is locked

Mode 005: Locked Access

- All buttons locked except the on/off button on battery powered
gauges.