5
Technical Data
Rotor FA-45-48-11
Rotor FA-45-20-17
Rotor FA-45-6-30
More technical data and order numbers can be found in the operating manual of Centrifuge
5804/5804 R/5810/5810 R.
Max. g-force (rcf)/speed (rpm):
Outer ring
Inner ring
19 083 × g/13 000 rpm
16 816 × g/13 000 rpm
Max. load per bore:
(Adapter, tube and contents)
3.75 g
Max. centrifugation radius:
Outer ring
Inner ring
10.1 cm
8.9 cm
Max. g-force (rcf)/speed (rpm):
20 913 × g/13 100 rpm
Max. load per bore:
(Adapter, tube and contents)
9.5 g
Max. centrifugation radius:
10.9 cm
Do not use Corning
®
50 mL PET Centrifuge Tubes. After centrifugation, the tubes
may get stuck in the rotor bores.
Max. g-force (rcf)/speed (rpm):
without adapter
with adapter for conical tubes 15 mL
5804/5804 R/5810
16 639 × g/11 000 rpm
16 233 × g/11 000 rpm
5810 R
20 133 × g/12 100 rpm
19 642 × g/12 100 rpm
Max. load per bore:
(Adapter, tube and contents)
75 g
Max. centrifugation radius:
without adapter
with adapter for conical tubes 15 mL
12.3 cm
12.0 cm