beautypg.com

Eppendorf C-76 Classic Water Bath Shaker User Manual

Page 33

background image

33

New Brunswick

Operating manual

8. Using your fingers, reach under the

SUBPLATFORM

and move the belt so that it falls

off the drive pulley. Remove the belt from the bearing housing pulley.

Figure 15: Belt Removal


9. Grab the belt from underneath the

SUBPLATFORM

, pull it forward and over one

corner of the

SUBPLATFORM

. Repeat for the other corner, until the belt clears the

SUBPLATFORM

, but is constrained by the

LOWER MOVER PLATE

.

10. Slide the belt to one of the unattached sides of the

LOWER MOVER PLATE

and pull it

through the opening.

11. Slide the belt to the other side of the

LOWER MOVER PLATE

, pull it through the

opening, and remove it from the unit.

12. Install the replacement belt in similar fashion.

13. Reattach the

UPPER FLEXURES

to the

LOWER MOVER PLATE

.

BEARING

HOUSING

PULLEY

DRIVE

PULLEY

BELT

LOWER
MOVER

PLATE