The Set and Maint indicators will flash and the accumulated running time will
be displayed in hundreds of hours (i.e., “02” equals 200 hours; “102” equals
10,200 hours). This display will continue for 10 seconds and then default to
the previous mode readout.

After 10,000 hours of operation, the Maint indicator will light. Preventive
maintenance is recommended at this point. The light can be deactivated by
Eppendorf service personnel.

NOTE:

Alteration of the internal clock by unauthorized personnel will void the
warranty.

3.5

Alarm Functions

Innova shakers have an audible alarm which is activated at predetermined times. It is
possible to deactivate and reactivate the alarm, according to your needs.

3.5.1

Stopping the Alarm

When the alarm sounds, you can stop it by pressing the

SELECT

key and

changing to any other function.

3.5.2

Deactivating the Alarm

The alarm may be deactivated in the following way:

1. Press

SELECT

to light HRS.

2. Simultaneously press the

σ and τ keys. The Set and Maint indicators will

flash.

3. While the Set and Maint indicators are flashing, press the

START/STOP

key. The Mute indicator will light to advise that the audible alarm is
deactivated.

3.5.3

Reactivating the Alarm

The alarm may be reactivated in the following way:

1. Press

SELECT

to light HRS.