beautypg.com

5 electrical connections, 6 platform installation, 7 clamp installation – Eppendorf 5000 Digital Tier Shake User Manual

Page 15: Lectrical, Onnections, Latform, Nstallation, Lamp

background image

15

Innova 5000

M1196-0050

Operating Manual

2.5

Electrical Connections

1. Check the voltage and frequency selector switches on the rear of the unit to ensure

that they are set to the appropriate voltage and frequency.

2. Remove the Caution label from the rear of the unit.
3. Set the circuit breaker on the front of the unit to the

OFF

position.


ONLY THEN:
4. Connect the power cord to the power cord connection, then plug the unit into a

grounded electrical outlet.

2.6

Platform Installation

A platform must be installed on the unit prior to use.

1. Place the platform on the shaker. Be sure to use the proper size platform for your

particular model shaker.

2. Tighten the platform locking levers.

2.7

Clamp Installation

Flask clamps purchased for use with universal platforms require installation.
Secure the base of the clamp to the platform with the correct type and number of
screws (refer to the clamp hardware application charts in Section 2.8).
All clamps are shipped complete with hardware.







CAUTION!

Before making electrical connections, be sure to
follow these instructions:

CAUTION!

A grounded electrical outlet is necessary for the safe
operation of this instrument.