9.5
Acceleration times and deceleration times for the Centrifuge 5427 R
(according to DIN 58 970)
Rotor
Acceleration time/Deceleration time
Mains voltage
230 V
120 V
100 V
FA-45-12-17
Without soft ramp Acceleration time
27 s
29 s
29 s
Deceleration time
24 s
25 s
25 s
With soft ramp
Acceleration time
39 s
39 s
39 s
Deceleration time
39 s
39 s
39 s
FA-45-24-11
Without soft ramp Acceleration time
18 s
19 s
19 s
Deceleration time
18 s
19 s
19 s
With soft ramp
Acceleration time
29 s
29 s
29 s
Deceleration time
31 s
31 s
31 s
FA-45-24-11-Kit
Without soft ramp Acceleration time
21 s
22 s
22 s
Deceleration time
21 s
21 s
21 s
With soft ramp
Acceleration time
32 s
32 s
32 s
Deceleration time
31 s
31 s
31 s
FA-45-30-11
Without soft ramp Acceleration time
21 s
22 s
22 s
Deceleration time
18 s
19 s
19 s
With soft ramp
Acceleration time
32 s
32 s
32 s
Deceleration time
33 s
33 s
33 s
F-45-30-11
Without soft ramp Acceleration time
21 s
22 s
22 s
Deceleration time
18 s
19 s
19 s
With soft ramp
Acceleration time
29 s
29 s
31 s
Deceleration time
32 s
32 s
32 s
FA-45-48-11
Without soft ramp Acceleration time
28 s
29 s
29 s
Deceleration time
22 s
23 s
23 s
With soft ramp
Acceleration time
36 s
36 s
39 s
Deceleration time
35 s
35 s
35 s
F-45-48-11
Without soft ramp Acceleration time
28 s
29 s
29 s
Deceleration time
22 s
23 s
23 s
With soft ramp
Acceleration time
36 s
36 s
36 s
Deceleration time
35 s
35 s
35 s
F-45-48-PCR
Without soft ramp Acceleration time
11 s
12 s
12 s
Deceleration time
12 s
13 s
13 s
With soft ramp
Acceleration time
22 s
22 s
22 s
Deceleration time
22 s
22 s
22 s
S-24-11-AT
Without soft ramp Acceleration time
18 s
18 s
18 s
Deceleration time
17 s
17 s
17 s
With soft ramp
Acceleration time
29 s
29 s
29 s
Deceleration time
30 s
30 s
30 s