5 aerosol-tight centrifugation, 6 safety instructions located on the device, Aerosol-tight centrifugation – Eppendorf Centrifuge 5427 R User Manual
Page 20: Safety instructions located on the device
Safety
Centrifuge 5427 R
English (EN)
20
3.5.5
Aerosol-tight centrifugation
3.6
Safety instructions located on the device
WARNING! Risk to health due to limited aerosol tightness with incorrect rotor/rotor lid
combination.
Aerosol-tight centrifugation is guaranteed only if the rotors and rotor lids intended for this
purpose are used. The designation of aerosol-tight fixed-angle rotors always starts with
FA.The aerosol-tight rotors and rotor lids of this centrifuge are additionally marked with a red
ring on the rotor and a red rotor lid screw.
Aerosol-tight swing-bucket rotors are marked with AT (aerosol-tight).
For aerosol-tight centrifugation, always simultaneously use rotors and rotor lids which are
marked as aerosol-tight in the centrifuge intended for the corresponding purpose. The
details specifying in which centrifuge you may use the aerosol-tight rotors and rotor lids
can be found on the rotor and, beginning from production date of October 2003, on the
upper side of the rotor lid.
Only use aerosol-tight rotor lids in combination with rotors which are marked on the rotor
lid.
WARNING! Damage to health as a result of limited aerosol tightness and incorrect usage.
Autoclaving, mechanical stresses and contamination by chemicals or other aggressive
solvents can impair the aerosol-tightness of the rotors and rotor lids.
Check the integrity of the seals of the aerosol-tight rotor lids or caps before each use.
Only use aerosol-tight rotor lids or caps if the seals are undamaged and clean.
Lightly grease the threads of the rotor lid screw with pivot grease after every proper
autoclaving (121 °C, 20 min.) (int. order no. Int. 5810 350.050, North America
022634330).
Replace aerosol-tight rotor lids and caps after 50autoclaving cycles.
For QuickLock rotor lids, the seal must be replaced after 50 autoclaving cycles.
Never store aerosol-tight rotors or buckets closed.
Display
Meaning
Location
Hazard point
Right side of the device
CAUTION
Always tighten the rotor using the supplied rotor
key.
Top of device, below the
centrifuge lid
CAUTION
Close all tubes and use a rotor lid.
Top of device, below the
centrifuge lid