Maintenance
Centrifuge 5427 R
English (EN)

38

6.2.1

Cleaning and disinfecting the device

1. Open the lid. Switch off the device with the mains/power switch. Disconnect the power plug from the

power supply.

2. Loosen the rotor nut by turning the rotor key counterclockwise.

3. Remove the rotor.

4. Clean and disinfect all accessible surfaces of the device, including the power cable, using a damp cloth

and the recommended cleaning agents.

5. Thoroughly clean the rubber seals of the rotor chamber with water.

6. Rub the dry rubber seals with glycerine or talcum powder to prevent them from becoming brittle. Other

components of the device, such as the lid latch, motor shaft and rotor cone, must not be lubricated.

7. Clean the motor shaft with a soft, dry and lint-free cloth. Do not lubricate the motor shaft.

8. Check the motor shaft for damage.

9. Inspect the device for corrosion and damage.

10. Leave the centrifuge lid open when the device is not being used.

11. Only connect the device to the power supply if it is fully dry inside and out.

6.2.2

Cleaning and disinfecting the rotor

1. Inspect the rotor and accessories for damage and corrosion. Do not use any damaged rotors or

accessories.

2. Clean and disinfect the rotors and accessories with the recommended cleaning agents.

3. Clean and disinfect the rotor lids. ONLY QuickLock: It is imperative that the sealing ring be removed

during this process so that the groove below it can be thoroughly cleaned.

Autoclaving
All rotors, rotor lids and adapters can be autoclaved (121 °C, 20 min).

Aerosol tightness
Check that the seals are intact before use.
Only QuickLock rotor lid: Replace the sealing ring in the lid groove when it becomes worn.
The sealing rings require regular care to protect the rotors.
Aerosol-tight rotors should never be stored with lids screwed on!
In order to prevent damage, lightly lubricate the lid thread of the aerosol-tight rotors with
pivot grease (order no. Int.: 5810 350.050/North America: 022634330).