Operation

Centrifuge 5427 R

English (EN)

25

5.2

Menu navigation

Proceed as follows to change settings in the device menu:

5.3

Menu

Tab. 5-1: Menu structure of the Centrifuge 5427 R.

1.

Open the menu.

2.

Select the desired menu item.

3.

Confirm your selection.

4.

Select the setting of the corresponding parameters.

5.

Confirm the changed setting.
The BACK menu item belonging to the first menu level appears.

6.

Exit the menu.

To exit the second menu level without changing a parameter, select the BACK menu item and
confirm with menu/enter.

Menu level 1 (M 1)

Menu level 2 (M 2)

Display

SOFT

Soft ramp: decreases acceleration speed
and braking speed
Disabled with short spin centrifugation

On

Rotor accelerates and brakes slowly.

OFF

Rotor accelerates and brakes rapidly.

RAD

For the internal conversion of speed (rpm)
to g-force (rcf), the radius is dependent on
the adapter used.

0_2ML
0_4ML
0_5ML
0_6ML

Select the radius according to the adapter
used.
Only rotor FA-45-12-17:

MAX
1_5/2
HPLC
CRYO

MAX

Largest radius of the used rotor.

LOCK

Key lock: centrifugation parameters
(temperature, g-force (rcf) or speed (rpm))
cannot be accidentally changed.

On

Set the centrifugation parameters
permanently.

SAFE

appears in the display

when the time, temp or speed keys are
pressed.

OFF

ATSET

Set start of centrifuging run time.

On

The set run time will be counted down
after 95 % of the specified g-force (rcf) or
speed (rpm) has been reached.

OFF

The set time is counted down immediately.