Centrifuge 5427 R
English (EN)

18

3.5.2

Incorrect handling of the centrifuge

3.5.3

Incorrect handling of the rotors

CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf may
impair the safety, functioning and precision of the device. Eppendorf cannot be held liable or
accept any liability for damage resulting from the use of incorrect or non-recommended
accessories and spare parts, or from the improper use of such equipment.

 Only use accessories and original spare parts recommended by Eppendorf.

NOTICE! Damage to device due to spilled liquids.

1. Switch off the device.
2. Disconnect the device from the power supply.
3. Carefully clean the device and the accessories in accordance with the cleaning and

disinfection instructions in the operating manual.

4. If a different cleaning and disinfecting method is to be used, contact Eppendorf AG to

ensure that the intended method will not damage the device.

NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool environment to a
warmer environment.

 After installing the device, wait for at least 4 h. Only then connect the device to the mains

power supply.

NOTICE! Damage from knocking against or moving the device during operation.
If the rotor bangs against the rotor chamber wall, it will cause considerable damage to the
device and rotor.

 Do not move or knock against the device during operation.

WARNING! Risk of injury from improperly attached rotors and rotor lids.

 Only centrifuge with rotor and rotor lid firmly tightened.
 If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be

properly secured. Immediately press the start/stop key to stop centrifuging.

CAUTION! Risk of injury due to asymmetric loading of a rotor.

 Load rotors symmetrically with identical tubes.
 Only load adapters with suitable tubes.
 Always use the same type of tubes (weight, material/density and volume).
 Check symmetric loading by balancing the adapters and tubes used with scales.