BNC 1120 - 1pS Time Interval Counter User Manual
Page 35

Model 1120 - 1pS Time Interval Counter
NUT 012
Ed. 6
Date : 18/01/2011
BERKELEY NUCLEONICS
Page
-35 / 36 -
6.1.3. Command structure
STAT
Syntax:
STAT CLEAR
STAT? POW|TRIG|CLK|TEMP|POPT|SYN
Form:
Set & Query
Description:
Clear or query the status.
Command
Parameter:
CLEAR: clear the trig status
Query
Parameter
SYN return the synthesis word
01 power fault
02 sync error
04 inhibited
POW return the power levels (+6V, -6V, +3.3V, +1.8V, +12V)
TRIG return the trigger status
CLK return the clock status
TEMP return the equipment temperature
POPT return the optical received power
Example:
STAT CLEAR
STAT? SYN => :STAT SYN,#H0001
STAT? POW => :STAT POW,6.00,-6.00,3.30,1.80,12.00
STAT? TRIG => :STAT TRIG,1,1
STAT? CLK => :STAT CLK,INTERNAL
STAT? POPT => :STAT POPT,4.54
STAT? TEMP => :STAT TEMP,45.27