BNC 1120 - 1pS Time Interval Counter User Manual
Page 23

Model 1120 - 1pS Time Interval Counter
NUT 012
Ed. 6
Date : 18/01/2011
BERKELEY NUCLEONICS
Page
-23 / 36 -
RUN
Syntax:
RUN DELAY|FREQ|COUNT
RUN?
Form:
Set & Query
Description:
Start measure or ask if measuring.
Command
Parameter:
DELAY start a delay measurement
FREQ start a frequency measurement
COUNT start a count measurement
Query response
:RUN YES measurement running
:RUN CAL calibration running
:RUN NO no measurement running
Example:
RUN DELAY
RUN? => :RUN YES
SAMPLES
Syntax:
SAMPLES
SAMPLES?
Form:
Set & Query
Description:
Set or query the measurement samples.
Command
Parameter:
Specified limit
1 <= samples <= 1000
Example:
SAMPLES 500
SAMPLES? => :SAMPLES 500