BNC 1120 - 1pS Time Interval Counter User Manual
Page 18

Model 1120 - 1pS Time Interval Counter
NUT 012
Ed. 6
Date : 18/01/2011
BERKELEY NUCLEONICS
Page
-18 / 36 -
CAL
Syntax:
CAL
Form:
Set
Description:
Start an auto calibration sequence.
Command
Parameter:
-
Query response
-
Example:
CAL
COUNT
Syntax:
COUNT?
Form:
Query
Description:
Query the counter or frequency measurement.
Query
Parameter:
-
Example:
COUNT? => :COUNT 32456123
DELAY
Syntax:
DELAY? [MEAN|MIN|MAX]
Form:
Query
Description:
Query the delay measurement in picoseconds.
Query
Parameter:
MEAN return the mean value of the samples
MIN return the minimum value of the samples
MAX return the maximum value of the samples
Example:
DELAY? MEAN => :DELAY MIN, 145
DELAY ? => :DELAY 141,145,148