BNC 1120 - 1pS Time Interval Counter User Manual
Page 20
Model 1120 - 1pS Time Interval Counter
NUT 012
Ed. 6
Date : 18/01/2011
BERKELEY NUCLEONICS
Page
-20 / 36 -
LEVEL
Syntax:
LEVEL START|STOP,
LEVEL? START|STOP
Form:
Set & Query
Description:
Set or query channel trigger level.
Command
Parameter:
START set level for START channel
STOP set level for STOP channel
Specified limit
0.5V <= level <= 5.0V
Example:
LEVEL START,2.8
LEVEL? START=> :LEVEL START,2.8
MODE
Syntax:
MODE SING|REP
MODE?
Form:
Set & Query
Description:
Set or query mode.
Command
Parameter:
SING set single measurement
REP set repetitive measurement
Query response
:MODE SING single measurement
:MODE REP repetitive measurement
Example:
MODE SING
MODE? => :MODE SING