BNC 835-M User Manual
Page 21
21
Model 835-M Programmer’s Manual V1.1 2011
:INITiate Subsystem
The :INITiate subsystem controls the state of the Model 835-M trigger system. The subsystem
commands and parameters are described below. The :INITiate commands, along with the
:ABORt and :TRIGger commands, comprise the Trigger Group of commands.
Command
Parameters
Unit
(default)
Remark
:INITiate[:IMMediate]
ON
V0.3.10
:INITiate:CONTinuous
ON|OFF|1|0
OFF
V0.3.10
:INITiate[:IMMediate]
Sets Model 835-M trigger to the armed state.
:INITiate:CONTinuous ON|OFF|1|0
Continuously rearms the Model 835-M trigger system after completion of a
triggered sweep.