BNC 835-M User Manual
Page 11
11
Model 835-M Programmer’s Manual V1.1 2011
The Service Request Enable (SRE) command sets the value of the Service Request Enable Register. The
variable is the decimal sum of the bits that will be enabled. Bit 6 (value 64) is ignored and
cannot be set by this command.
Range 0–255
The setting enabled by this command is not affected by signal generator preset or
*RST. However, cycling the signal generator power will reset it to zero.
*SRE?
The Service Request Enable (SRE) query returns the value of the Service Request Enable
Register.
Range 0–63 & 128-191
*STB?
The Read Status Byte (STB) query returns the value of the status byte including the master
summary status (MSS) bit.
Range 0–255
*TRG
The Trigger (TRG) command triggers the device if LAN is the selected trigger source, otherwise,
*TRG is ignored.
*TST?
The Self-Test (TST) query initiates the internal self- test and returns one of the following results:
0 This shows that all tests passed.
1 This shows that one or more tests failed.
*WAI
The Wait- to- Continue (WAI) command causes the signal generator to wait until all pending
commands are completed, before executing any other commands.