Document # 99300 RevA

8

Printed in USA © 2008

WARNING

Keep personnel and equipment clear of table be-

fore initiating movement to avoid personal injury

or damage to the equipment.

Foot Control (Figure 9)

Depress the pedal at the desired location to activate the

function shown.

NOTE: The table incorporates a safety switch which pre-

vents the leg rest from moving if the treatment pan is not

fully stored (inward).

CAUTION

The foot control is moisture resistant and runs on

low voltage, eliminating shock hazards. However,

to avoid temporary or permanent damage to the

foot control, do not immerse in water or liquids.

CAUTION

Keep foreign objects away from foot control. To

avoid accidental operation, be sure the foot control

is NOT positioned below the legrest or headrest

linkage. Failure to do so could result in unwanted

motion and possible damage to the table.

General

DANGER

Do not use this table in an explosive or oxygen-

rich atmosphere. To do so will result in an explo-

sion or fire.

WARNING

If the table malfunctions, immediately remove your

hand or foot from the controls, unplug the power

cord from the electrical receptacle, and assist the

patient from the table. If table continues to mal-

function, call for service. Do not attempt to repair

table yourself.

WARNING

Before initiating power, keep personnel and equip-

ment clear of the table movement to avoid person-

al injury or equipment damage.

OPERATION

NOTE: Each control function will automatically stop mov-

ing when either the UP or DOWN travel limit is reached.

If the button is pressed with travel at its limit, the actua-

tor will not run. Sensors prevent the actuator motor from

operating when travel limit has been reached, preventing

wear of the actuator.

NOTE: For optimum table performance, allow the table to

reach room temperature before operating.

NOTE: The table is not designed for continuous opera-

tion. The control system includes a thermal overload

safety switch to protect from overheating. If the table is

operated continuously causing the control to exceed its

allowable operating temperature, the thermal overload

safety switch will shut off. If normal operation ceases, do

not attempt to operate the table. Allow the table to cool

for 10-15 minutes before attempting to use again. The

thermal overload switch will automatically reset when it

has reached a safe operating temperature.

NOTE: The control system is protected by a fuse to

prevent excessive current draw which could damage

the system. If operation does not resume after a cooling

period of one hour, contact your dealer or an authorized

service center for repair.