beautypg.com

Warning – Brewer Assist Power User Manual

Page 14

background image

Document # 99300 RevA

14

Printed in USA © 2008

Figure 21. Treatment Pan

Figure 22. Headrest Handles

Treatment Pan (Figure 21)

NOTE: A safety switch prevents operation of the foot

function when the treatment pan tray is extended. This

prevents the legrest from contacting the treatment pan

tray. If the legrest is inoperable, check to ensure that the

treatment pan tray is in its fully stowed position (inward).

A light clicking sound will be heard just before the treat-

ment pan reaches its fully stowed position.

1. To expose the treatment pan for use, grasp the tray

handle and pull until tray is fully extended (Figure 21).

2. The pan can be easily removed for cleaning with

the tray fully extended. Fully seat pan into tray when

clean.

3. Push the tray handle as far in as it will go to fully stow

the treatment pan after use. A light clicking sound will

be heard just before the pan reaches its fully stowed

position.

Headrest Positions (Figure 22)

WARNING

Make sure headrest is securely locked into position

before use. Failure to have the three handles cor-

rectly locked could result in headrest falling during

a procedure which could result in serious personal

injury.

1. To move the headrest, rotate the three handles to

their open position. Move the headrest to its desired

location. Rotate the 3 handles to their locked position

as shown in Figure 22.

2. Each section of the headrest can be adjusted inde-

pendently by rotating the handle to the open position,

moving the desired section, and then rotating the

handle to the closed position.

NOTE: Make sure all three handles are locked. A handle

is locked correctly when it is pushed past center and

the force of the clamping action pulls the handle into the

locked position and holds it there.

WARNING

If any of the handles fails to lock properly, the

headrest assembly could fall during a procedure.

This could cause severe injury to a patient. Do not

attempt to repair the headrest assembly yourself;

call for service.

TABLE ADJUSTMENTS (CONTINUED)

Headrest Handle

(Locked)

Upper Handle

(Locked)

Lower Handle

(Locked)

Tray

Treatment

Pan