Calibration tab

Figure 44: Calibration tab

Here you can calibrate display(s), revert to the latest calibration,
revert to default factory settings and view the result of the latest
calibration. The possible options are explained below.

To calibrate the display using a sensor:

1. If you have multiple displays of the same type connected

(multi-head system), consider if you wish to equalize the
display you calibrate to a reference display.

Equalization means the luminance of the display you are
calibrating will be matched to the luminance of the reference
display.

If you wish to equalize the display, you must close the
Display Settings dialog and make sure the Equalization