CHAPTER 4
Using terminal commands
33
APPLIES
TO
A720, A720B, A723
D
ESCRIPTION
Sets/returns the input storing and sampling intervals.
P
ARAMETERS
storage represents the time (in seconds) elapsed between two
slots stored in the internal memory, while samples represents the
numbers of samples used to build the average that will be stored.
R
ETURNS
The interval and rate.
REMARKS
The default storage is 900 (15 minutes) and samples is 3 (3 sam-
ples per quarter of an hour).
REMOTE
Yes, SET only.
E
XAMPLE
SLOT 900 3
193 SLOT 0
#
SLOT
193 SLOT 900 3 0
#
Note: The A720B base station needs only the rate. For consis-
tency, however, it accepts both parameters and discards the
one it doesn’t need.
C
OMMAND
PMP
APPLIES
TO
A720, A720B, A723
D
ESCRIPTION
Sets/returns the node’s Power Management Parameters (switches
on/off the battery charge).
P
ARAMETERS
The lower (switch on) and the higher limit (switch off), both in volts
x 10. Standard Values are 65 (for 6.5 Volts) for switch on and 72 (for
7.2 Volts) for switch off.
R
ETURNS
The lower (switch off) and the higher limit (switch on), both in volts
x 10.
REMARKS
GET/SET.
REMOTE
Yes, SET only.
E
XAMPLE
PMP 65 72
193 PMP 0
#