perfect accuracy. Incorrect calibration can occur if you do not follow the steps exactly. You will need a
50g weight (DURA 50) /100g weight (DURA 100) to calibrate this scale.
1) Turn the scale OFF and place it on a FLAT, very stable surface .
2) Press and holdUNIT andON/OFF to turn on the scale, wait for the display to show “CALE”.
3) Place the MAX.CAPACITY weight (50g for DURA 50, 100g for DURA 100) on the tray, wait 3 seconds
then press TARE. Calibration is complete.
NOTE: If after calibration your scale does not read accurately, this indicates calibration error and the
calibration process should be repeated more slowly.
Please calibrate on a very stable flat surface!
INACCURACY/ERROR
The primary reasons for inaccuracy or malfunction are low batteries, incorrect calibration, overload or
operating on an unstable surface. Please keep this in mind and maintain and operate your scale
properly. The scale is a precise instrument and must be handled with the utmost care and caution.
FEATURE
Ø Power Up Segment Test
When first turning the unit on, all segments of the display will appear from“99999” to“00000”, and all
the indicator will light.
Ø Overload
When an applied load exceeds the maximum capacity. It will display “- - - - -“, & “OVER LOAD”, Remove
the excessive load immediately!
Remember: You can permanently damage the scale by overloading it!
Ø Negative Value
Any tared value or a value left in memory will be displayed as a negative number once all weight is
removed from the tray. Press TARE to re-Zero the unit. It will show “_ _ _ _ _” when the weight is below
than the range to display.
Note: “-.****” means “- 0.****”