Ids industrial dispenser instructions – FMI IDS User Manual
Page 10
IDS INDUSTRIAL DISPENSER
INSTRUCTIONS
IN-IDS-08
10
1
IDS-“Q” Pumphead Displacement is adjusted by CALIBRATION RING FIXED CALIBRATION LINE The Knurled ADJUSTMENT NUT on the IDS-“RH” Pumphead 7.0 Calibration and Displacement Your IDS comes factory calibrated to maximum displacement of the installed pumphead unless ordered 7.1 Calibration for Dispensing
direction
changing Angle of pumphead.
Loosen 11/32 hex nut and move
pumphead “left to increase” and
“right to decrease”.
Now recalibrate. Lock hex nut set
and lock Displacement adjust
stop screw.
pumphead controls stroke to stroke
Piston displacement. Turning clockwise
to zero stops displacement. Turning the ADJUST-
MENT-NUT counterclockwise four and one half turns
from zero (450 on scale) (fig.5) causes maximum
pump reciprocation, e.g., 50 µl per stroke for the H-0
or 100 µ for H-1 unit.
Thus each 1-1/8 turn (112.5 on scale) of the AD-
JUSTMENT NUT Represents 25% of maximum
(12.5µl for H-0 and 25µl for H-1) and each graduation
on the ring represents an adjustment of 1/450th of
maximum (0.111 µl for H-0, 0.222 µl for H-1).
Flow direction> left to right
with specific displacement and calibration.
Recalibration and changes of displacement are relatively easy.
FMI calibrates by weighing 3 or more dispenses on an analytical balance. It then finds the average. The
pumphead displacement is then adjusted as required up or down accordingly. See Fig