beautypg.com

7 syringe diameters and rate limits – Next Advance SP2200 User Manual

Page 51

background image

NEXT>>>ADVANCE

www.nextadvance.com

SP2200Programmable Syringe

Pump

Publication #1200-02

46

04/29/10

10.7 Syringe Diameters and Rate Limits

Syringe

Manufacturer

Syringe

(cc)

Inside

Diameter

(mm)

Maximum

Rate

(mL/hr)

Minimum

Rate

(µl/hr)

Maximum

Rate

(mL/min)

B-D

1

4.699

191.1

1.459

3.185

3

8.585

637.9

4.868

10.63

5

11.99

1244

9.495

20.74

10

14.43

1802

13.76

30.04

20

19.05

3141

23.97

52.35

30

21.59

4035

30.79

67.25

60

26.59

6120

46.7

102

HSW

1

4.69

190.4

1.453

3.173

Norm-Ject

3

9.65

806.1

6.151

13.43

5

12.45

1341

10.24

22.36

10

15.9

2188

16.7

36.47

20

20.05

3479

26.55

57.99

30

22.9

4539

34.64

75.65

50

29.2

7380

56.32

123

Monoject

1

5.74

285.2

2.176

4.753

3

8.941

692

5.28

11.53

6

12.7

1396

10.66

23.26

12

15.72

2139

16.33

35.65

20

20.12

3504

26.74

58.4

35

23.52

4788

36.54

79.81

60

26.64

6143

46.88

102.3

140

38

95.37

208.3

Terumo

1

4.7

191.2

1.459

3.187

3

8.95

693.4

5.291

11.55

5

13

1462

11.17

24.38

10

15.8

2160

16.49

36.01

20

20.15

3514

26.82

58.57

30

23.1

4619

35.25

76.98

60

29.7

7635

58.26

127.2

Air-Tite

10

15.9

2188

16.7

36.47

20

20.25

3549

27.09

59.16

30

22.5

4382

33.44

73.03

50

29

7280

55.55

121.3

Syringe

(uL)

Inside

Diameter

(mm)

Maximum

Rate

(ul/hr)

Minimum

Rate

(µl/hr)

Hamilton

0.5

0.103

91.83

0.001

Microliter

1

0.146

184.5

0.002

2

0.206

367.3

0.003

5

0.326

919.9

0.008

SGE

0.5

0.1

86.56

0.001

1

0.15

194.7

0.002