2 rs-232 pump network connector wiring – Next Advance SP2200 User Manual
Page 47
NEXT>>>ADVANCE
www.nextadvance.com
SP2200Programmable Syringe
Pump
Publication #1200-02
42
04/29/10
*ADR [ DUAL | RECP ]
Set Reciprocating or Dual pumping mode
*RESET
Clears program memory and resets communication parameters
(system command, valid regardless of current address)
------------------------------------------------------------------------------------------------------------------------------
Network Command Burst =>
Communicate with multiple pumps on a pump network simultaneously
10.2 RS-232 Pump Network Connector Wiring
RS-232
To Network To Computer
PC Com Port Connectors
25 Pin
9 Pin
3 - Receive
2 - Receive
Connect to pump TXD
2 - Transmit
3 - Transmit
Connect to pump RCV
9 - Ground
5 - Ground
Connect to pump GND
10.3 Accessories
10.3.1 ANA-BOX
Part #: ADPT-ANABOX, Analog voltage control interface.
Allows the pump to be controlled by a variable voltage source. Allows the Creation of a closed loop system
allowing the NE-1000 to be controlled by an external sensor, such as a pressure sensor, or other variable
voltage source. Start or stop the pump at a specific voltage level. Set the pumping rate to be proportional to
the voltage input.
10.3.2 Syringe Heater
Part#: SYRINGE-HEATER, see www.SyringeHeater.com for details.
Flexible heating pad that wraps around the syringe. Digital PID controller will heat a syringe to a set
temperature up to 100 C.
10.3.3 RS-232 Network Cables
RS-232 Network Primary Cable
7 foot cable, part #:
CBL-PC-PUMP-7
25 foot cable, part #:
CBL-PC-PUMP-25
RCV from network
TXD to network
TXD to computer
RCV from computer