Starting with the second dispense, 0.25 is added to the volume dispensed to compensate for the sucked back
volume of the previous dispense. By changing the last Phase to a [FUN LOP nn] function, the total number of
dispenses can be set.
When entering a function with associated data, such as with the ‘Pause’ in Phase 5 or the ‘Loop’ in Phase 6, the
function is entered in 2 steps. First select the function and store it. Then enter the associated data.
Phase
Function
Rate
Volume
Direction
1
RAT
750 ml/hr
2.0 ml
Infuse
Phase
Function
Rate
Volume
Direction
2
RAT
750 ml/hr
0.25 ml
Withdraw
Phase
Function
3
LPS
Phase
Function
4
LPS
Phase
Function
5
PAS 90
Phase
Function
6
LOP 03
Phase
Function
7
BEP
Phase
Function
8
PS:30
Phase
Function
Rate
Volume
Direction
9
RAT
750 ml/hr
2.25 ml
Infuse
Phase
Function
Rate
Volume
Direction
10
RAT
750 ml/hr
0.25 ml
Withdraw
Phase
Function
11
LPE
7.4.3 Example 3: Ramping the Flow Rate
m
l/h
r
R
at
e
Time
0
150
200
250
Phases
2 - 4
Phases
5 - 8
Phases
9-11
Note: Graph is representative only.
Pumping rate increments and decrements in steps.
Continuously ramp up and down the pumping rate. Starting at 200 ml/hr, the pumping rate will increment to
250 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Then the pumping rate will decrement to
150 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Finally, the pumping rate is incremented
back to 200 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed, then the process is repeated.