beautypg.com

WIKA PGT15 User Manual

Page 10

background image

11412101 09/2009 GB/D/F/E

WIKA Operating instructions intelliGAUGE Model PGT15 (LOG-1x)

10

GB

9. Trouble shooting / 10. Technical data

Malfunction

Possible cause

Measure

No output signal

Failure of power supply

or wiring interrupted

Check power supply and

wiring; replace defective

components

Electronic defect

through incorrect

supply voltage or

external voltage

Return instrument to

manufacturer for repair

Steady signal despite

pressure changes

Pressure entry blocked Check tailpipes and

pressure entry bore, if ne-

cessary clean it carefully

Excessively high,

steady signal despite

pressure changes

Electronic defect

through incorrect

supply voltage or

external voltage

Return instrument to

manufacturer for repair

Full span reading

too low

Supply voltage too low Adjust supply voltage
Load impedance

too high

Consider max. permissi-

ble load

Zero point shifted

Return instrument to

manufacturer for repair

Zero signal too low /

troo high

Zero point shifted

Return instrument to

manufacturer for repair

10. Technical data

For technical data please refer to the technical data sheet.

PV 11.04 Model PGT15 (LOG-1x),

intelliGAUGE, Bourdon Tube Pressure Gauges with

Stepped Electrical Output Signal, NS 50