WIKA operating instructions intelliGAUGE / switchGAUGE

5

GB

3. Installation instructions

3. Installation instructions

Installation

Nominal position acc. EN 873-3 / 9.6.6 figure 7: 0° or 90°

Pressure connector bottom or rearside

Permissible vibration load at the mounting point

The instrument should always be installed in locations free of vibration.

If necessary, it is possible to isolate the instrument from the mounting

point by installing a flexible connection line between the measuring

point and the pressure gauge and mounting the instrument on a suitable

bracket.

If this is not possible, the following limit values must not be exceeded:

Unfilled devices:

Frequency range 10 … 150 Hz

Accelaration

< 0.5 g (5.0 m/s²)

Liquid filled devices:

Frequency range 10 … 150 Hz

Accelaration

< 4.0 g (39.2 m/s²)

The liquid filling has to be checked periodically.

The liquid level must not drop below 75 % of the instrument´s diameter.
Test connector

At certain use cases (e.g. steam vessels) isolating devices have to have a

test connector, so that the instrument can be checked without dismounting.
Mounting provisions

If the line for the pressure gauge is not rigid enough for vibration-free instal-

lation, fasten the instrument using appropriate fastening elements for wall

and / or pipe mounting, and, if necessary, by means of a capillary line.
Measuring system damping

If it is not possible to avoid vibration by means of appropriate mounting, use

pressure gauges with liquid filling.
Effects of temperature

The mounting of the pressure instrument has to be done in a way that the

operating temperature is neither exceeded nor falls below the permissible

operating temperature, even in consideration of thermal convection and

thermal radiation. Tailpipes of a suitable length or syphons may be used to

protect the pressure instrument and the isolating device.

The influence of the temperature onto the accuracy of indication and of

measurement has to be noticed.