WIKA operating instructions DELTA-plus, DELTA-comb and DELTA-switch

21

14093265.01 03/2014 GB/D/F/E

7. Options and accessories

Specifications for handling

Sequence of operations to start measurement

1. Open the pressure compensating valve (middle valve spindle)

2. Open the shut-off valve for the negative media chamber (⊖, right-hand valve)

and the positive media chamber (⊕, left-hand valve)

3. Close the pressure compensating valve

Sequence of operations to flush/vent the measuring lines

1. Start: Open the shut-off valve for the ⊕ and ⊖ media chamber, open the

pressure compensating valve and vent valve

2. Finish: Close the pressure compensating valve and vent valve

Sequence of operations to finish measurement (also temporary shutdown)

1. Open the pressure compensating valve

2. Close the shut-off valve for the ⊕ and ⊖ media chamber

Sequence of operations to dismount the measuring instrument with a running

process

1. Close the shut-off valve for the ⊕ and ⊖ media chamber

2. Open the vent valve