Voltages greater than AC 50 V or DC 120 V:

- Do not connect circuits simultaneously with extra-low voltage circuits or with

safety extra-low voltage (SELV) or protected extra-low voltage (PELV).

- Circuits must offer a device, external to the measuring instrument, that

enables the instrument to be isolated from the electrical supply. This must be

easily accessible and be marked as the isolation device for the instrument.

- Cables for the circuit must fulfil the isolation requirements and conform to, for

example, IEC 60227 or IEC 60245.

With flexible connecting cables, use isolated end splices.

Connecting cables must be suited to the ambient temperature range of the

application.

Seal the cable entry with the appropriate approved cable glands.

6. Commissioning, operation

Only use cable

with a diameter

of 7 ... 13 mm

Cable gland design

1346210y

Install the connection cables securely.