beautypg.com

Options and accessories – WIKA 851.02.100 User Manual

Page 9

background image

WIKA Operating instructions DELTA-plus, DELTA-comb and DELTA-switch

9

308

15 07/

009 GB/D/F

GB

7. Options and accessories

7.2 Four-way valve manifold

Shut off j- and i process line for

disassembling and testing of

measuring instrument without interrupting the process that is running.


Protect the instrument against inadmissible

overpressure loading, such as, for example, in

the case of pressure tests and undefined
operating conditions (also temporary shut down).

Pressure equalising for

zero point check during

normal operation as well as equalisation of the

pressures at high pressure and low pressure

side during start-up and operation (with opened
pressure equalising valve).

Process line

air bleeding for liquid measuring

media and

purging the process line for removing

impurities.

7.2.1 Handling instructions

Work sequence to

start measuring

1. Open pressure equalising valve (middle valve)
. Open shut-off valve of the negative-pressure media chamber (i right

valve) and the positive-pressure media chamber (j left valve)
3. Close pressure equalising valve

Operating sequence for

air bleeding/purging of piping

1. start: Open shut-off valve of the negative-pressure (i) and positive-

pressure (j) media chamber

Open pressure equalising valve and valve for purging or air bleeding

. end: Close pressure equalising valve and valve for purging or air bleeding

Procedure at the

end of the measuring operation

(also for partial system shut-down)

1. Open pressure equalising valve
. Close shut-off valve of the negative-pressure (i) and positive-pressure

(j) media chamber

Task list for

removing the transmitter during normal operation

1. Open pressure equalising valve
. Close shut-off valve of the negative-pressure (i) and positive-pressure

(j) media chamber
3.

Open valve for purging or air bleeding

Gauge valve,
low pressure
side i

Pressure
equalising valve

Gauge valve,
high pressure
side j

Valve for
purging or
air bleeding

This manual is related to the following products: