9. Dismounting, return and disposal

WARNING!

Residual media in dismounted pressure switches can result in a risk to persons, the

environment and equipment.

Take sufficient precautionary measures.

9.1 Dismounting

Only disconnect the pressure switch once the system has been depressurised!

9. Dismounting, return and disposal

Problem

Possible cause

Measure

Signal span drops

Humidity has entered

Assemble the cable correctly

Signal span drops/too small

Diaphragm damaged, e.g. due to

impacts, abrasive/aggressive medium;

corrosion at diaphragm/process

connection

Contact manufacturer and replace

instrument

If complaint is unjustified, we will charge you the complaint processing fees.

CAUTION!

If deficiencies cannot be eliminated by means of the measures listed above, shut down the

instrument immediately, and ensure that pressure and/or signal are no longer present, and

secure the instrument from being put back into operation inadvertently. In this case, contact

the manufacturer. If a return is needed, follow the instructions given in chapter 9.2 "Return".